NETZSCH - Leading Thermal Analysis.    
   
 
 
 

Accurate thermal conductivity testing of insulation materials

HFM 436 Lambda - Heat flow meter

 
GeneralTechniqueSoftwareApplicationsAccessoriesLiterature 

HFM 436 Lambda - Technical Specification (subject to change)

  HFM 436/3/0 HFM 436/3/1 (E) HFM 436/6/1E
Plate Temperature Ranges: 0°C to 40°C fixed 0°C to 100°C (-30°C to 90°C at HFM 436/3/1E) variable -20°C to 70°C variable
Cooling System: Forced Air Forced Air (at /1E external chiller) External chiller
Plate Temperature Control: Peltier System Peltier System Peltier System
Programmable Data Points: 1 10 10
Specimen Size: 30x30x10 cm³ 30x30x10 cm³ 60x60x20 cm³
Thermal Resistance Range: 0.1 to 8.0 m²K/W 0.1 to 8.0 m²K/W 0.1 to 8.0 m²K/W
Thermal Conductivity Range: 0.005 to 0.50 W/mK
Repeatability: 0.5 % 0.5 % 0.5 %
Accuracy: + / - 1 to 3 %
Dimensions (LxWxH): 48x63x51 cm³ 48x63x51 cm³ 80x95x80 cm³


Schematic of the HFM 436Lambda, principle of operation

 

GeneralTechniqueSoftwareApplicationsAccessoriesLiterature 

Please also visit the following product pages:

LFA 447 NanoFlash®

LFA 447 NanoFlash®


 
    print this page back to top back  
 

Thermal Analysis News | Thermal Analysis Products | Thermal Analysis Market Segments | Materials Characterization | Service for Thermal Analysis | Seminars for Thermal Analysis | Applications Measurements | Compliance | Applications Advisory Service | Literature Data Base | Proteus Software | Advanced Software | Contact | NETZSCH worldwide | History | DSC Analysis | TG Analysis | TG-DSC Analysis (STA) | QMS / FTIR Analysis | DIL / TMA / DMA | DEA | LFA / HFM / TCT | RUL / HMOR / PCE / TCT | Sitemap | Imprint