| | HFM 436/3/0 | HFM 436/3/1 (E) | HFM 436/6/1E |
| Plate Temperature Ranges: | 0°C to 40°C fixed | 0°C to 100°C (-30°C to 90°C at HFM 436/3/1E) variable | -20°C to 70°C variable |
| Cooling System: | Forced Air | Forced Air (at /1E external chiller) | External chiller |
| Plate Temperature Control: | Peltier System | Peltier System | Peltier System |
| Programmable Data Points: | 1 | 10 | 10 |
| Specimen Size: | 30x30x10 cm³ | 30x30x10 cm³ | 60x60x20 cm³ |
| Thermal Resistance Range: | 0.1 to 8.0 m²K/W | 0.1 to 8.0 m²K/W | 0.1 to 8.0 m²K/W |
| Thermal Conductivity Range: | 0.005 to 0.50 W/mK |
| Repeatability: | 0.5 % | 0.5 % | 0.5 % |
| Accuracy: | + / - 1 to 3 % |
| Dimensions (LxWxH): | 48x63x51 cm³ | 48x63x51 cm³ | 80x95x80 cm³ |