Lead chloride (7.92 mg) in an argon flow of 150 ml/min shows evaporation starting in the melting range (487°C). The molecule ion (PbCl2 m/z = 278) and fragment ions caused by dissociation and ionization (PbCl m/z = 243, Pb m/z = 208, Cl m/z = 37, CL m/z = 35) are clearly detected far below the boiling temperature of the starting material.

Pitch carbon powder (55.2 mg) decomposes in nitrogen flow (50 ml/min) into aromatic compounds of high molecular weight, mainly up to 600°C. Only selection is shown by MID curves for pyrenes (m/z = 202), Triphenylenes (m/z = 228), Benzo(a)pyrenes (m/z = 252), Benzo(ghi)perylenes (m/z = 276) and Dibenzopyrenes (m/z = 302).

 

GeneraleTecnicaSoftwareApplicazioniAccessoriLetteratura 

Per favore, visitate anche le seguenti pagine prodotti:

STA 409 CD

STA 409 CD


STA 429 CD

STA 429 CD


 
    stampa la pagina torna su indietro  
 

Notizie | Prodotti | Rami / Industrie | Caratteristici dei materiali | Assistenza Tecnica | Seminari | Analisi e Applicazioni | Conformità | Servizio Applicazioni Suggerimenti | Letteratura | Software Proteus® | Advanced Software | Contatti | Locazioni | NETZSCH-storia | DSC/DTA | TG/TGA | STA (TG-DSC) | QMS / FTIR | DIL / TMA | DMA/DMTA | DEA | LFA/HFM/TCT | RUL/HMOR/PCE/TCT | Sitemap | Editoriale