Unique, integrated coupling system for unlimited gas analysisQMS 403/5 SKIMMER - System sprzężony SKIMMER
Info & service :Informacja o produkcie Informacja serwisowa Lead chloride (7.92 mg) in an argon flow of 150 ml/min shows evaporation starting in the melting range (487°C). The molecule ion (PbCl2 m/z = 278) and fragment ions caused by dissociation and ionization (PbCl m/z = 243, Pb m/z = 208, Cl m/z = 37, CL m/z = 35) are clearly detected far below the boiling temperature of the starting material. Pitch carbon powder (55.2 mg) decomposes in nitrogen flow (50 ml/min) into aromatic compounds of high molecular weight, mainly up to 600°C. Only selection is shown by MID curves for pyrenes (m/z = 202), Triphenylenes (m/z = 228), Benzo(a)pyrenes (m/z = 252), Benzo(ghi)perylenes (m/z = 276) and Dibenzopyrenes (m/z = 302).
Odwiedź następujące strony o naszych produktach
|
|||||||||||||||||||
|
|
||||||||||||||||||||
Analiza termiczna - Aktualności | Analiza Termiczna - Produkty | Segmenty rynku / przemysłu - Analiza termiczna | Materiały / Aplikacje - Analiza termiczna | Serwis - Analiza termiczna | Seminarium - Analiza termiczna | Applications Measurements | Compliance | Applications Advisory Service | Media Service | Proteus Software | Advanced Software | Kontakt | NETZSCH worldwide | History | DSC / DTA | Multiple Mode Calorimetry (MMC) | TGA Analysis | TG-DSC Analysis (STA) | QMS / FTIR, GC-MS | DIL | TMA | DMA / DMTA | Adiabatic Reaction Calorimetry | DEA | LFA / GHP / HFM / TCT | RUL / HMOR / PCE / TCT | Sitemap | Imprint Grupa NETZSCH | NETZSCH Grinding & Dispersing | NETZSCH Pumps |
||||||||||||||||||||