NETZSCH - Leading Thermal Analysis.   Leading Thermal Analysis  
   
 
 
 

Adiabatic Reaction Calorimetry

Accelerating Rate Calorimeter 254 (ARC®)

 
GeneralTechniqueApplicationsAccessoriesLiteratureVideos/Webcast 

Accelerating Rate Calorimeter 254 - Technical Specifications (subject to change)

Info & Service:

Product Information
Would you like to request more information on a particular product?
We will be happy to send you what you need.
Continue

Service Information
Find out more about our various service offerings.
We are at your assistance!
Continue

  • Minimum Detectable Slope: < 0.01 K/min
  • Operating Test Cell Temperature Range: RT to 500°C (sub-ambient optional)
  • Operating Test Cell Pressure Range: 0 to 200 bar
  • Pressure Accuracy: 0.1% of full scale
  • Sample size: 0.5 ml–7 ml
  • Tracking rate at adiabatic operation up to 200 K/min
  • Test cell materials:
    Hastelloy- C,
    titanium,
    tantulum,
    stainless steel,
    inconel
  • Temperature Accuracy: +/-0.1°C plus thermocouple uncertainty
  • Stirring, injection, and vent capability (options)
  • ASTM standard: E1981

 

GeneralTechniqueApplicationsAccessoriesLiteratureVideos/Webcast 
 
    print this page back to top back  
 

Thermal Analysis News | Thermal Analysis Products | Thermal Analysis Market Segments | Materials Characterization | Service for Thermal Analysis | Seminars for Thermal Analysis | Applications Measurements | Compliance | Applications Advisory Service | Media Service | Proteus Software | Advanced Software | Contact | NETZSCH worldwide | History | DSC Analysis | Multiple Mode Calorimetry (MMC) | TGA Analysis | TGA-DSC Analysis (STA) | QMS / FTIR, TG-GC-MS Analysis | DIL | TMA | DMA / DMTA | Adiabatic Reaction Calorimetry | DEA | LFA / GHP / HFM / TCT | RUL / HMOR / PCE / TCT | Sitemap | Imprint