DSC 204 F1 Phoenix®

The premium DSC for all requirements

The versatile Premium Differential Scanning Calorimeter, DSC 204 F1 Phoenix®, fascinates with its unique concept: all essential operational components are integrated into the instrument and optional add-ons can be installed anytime on site.

The DSC measuring cell consists of a cylindrical high-conductivity silver block with an embedded heating coil for broad thermal symmetry (3D symmetry) in the sample chamber, the cooling ports for liquid nitrogen or compressed air cooling and a cooling ring for connection of the intracooler (also with simultaneous liquid nitrogen cooling). The gas-tight construction and integrated mass flow controllers for the purge and protective gases allow coupling to an FT-IR or MS for gas analysis.

Tau-sensorTau-sensor

Exchangeable sensors are tailored to their tasks:

With its disk-shaped silver carrier plate and supersensitive thermosensors of nickel-chromium constantan, the Tau-sensor offers a high level of calorimetric sensitivity along with extremely short signal time constants of only 0.6 seconds, which guarantees good separation of overlapping thermal effects.

The µ-sensor stands out for its high level of calorimetric sensitivity, never before achieved with a Differential Scanning Calorimeter, DSC. For example, it is extremely well suited for pharmaceutical applications with small sample weights.

µ-sensorµ-sensor

More on the Functional Principle of a Heat-Flux DSC


Key Technical Data

(subject to change)

  • Temperature range:
    -180°C to 700°C
  • Heating rates:
    0.001 K/min to 200 K/min
  • Cooling rates:
    max. 200 K/min
  • Compressed air cooling:
    RT to 700°C
  • Intracooler for the extended range:
    -85°C to 600°C
  • Automatically controlled LN2 cooling:
    -180°C to 700°C
  • Indium Response Ratio (*):
    >/100 mW/K
    (*) Peak Height [mW]/Peak with at 50% Height [K] of Indium melting peak
  • Gas control for 2 sample gases and 1 protective gas through integrated mass flow controller and software
  • Automatic sample changer:
    for 192 samples and reference crucibles (optional)

Software

The DSC 204 F1 Phoenix® runs under Proteus® Software on Windows®. The Proteus® Software includes everything you need to carry out a measurement and evaluate the resulting data. Through the combination of easy-to-understand menus and automated routines, a tool has been created that is extremely user-friendly and, at the same time, allows sophisticated analysis. The Proteus® software is licensed with the instrument and can of course be installed on other computer systems.

DSC features:

  • Determination of onset, peak, inflection and end temperatures   
  • Automatic peak search   
  • Transformation enthalpies: analysis of peak areas (enthalpies) with selectable baseline and partial peak area analysis   
  • Evaluation of crystallization   
  • Comprehensive glass transition analysis   
  • Automatic baseline correction   
  • Specific heat determination   
  • BeFlat® for optimized baselines through a multi-dimensional, temperature and heating rate dependent polynomial   
  • Tau-R® Mode: takes into account the time constant and thermal resistance of the instrument and reveals thus sharper DSC effects from the sample   
  • Optional TM-DSC (temperature modulation) for separation of reversing (thermodynamic) and non-reversing (kinetic) effects   
  • Sector-specific software solutions for conformance to e.g. GLP and GMP

You can use the following software with this product:
Proteus® Software, Advanced Software


Accessories

The DSC 204 F1 Phoenix® is set up for various cooling systems. The entire temperature range from -180°C to 700°C can be covered by different liquid nitrogen cooling devices for LN2 and GN2 operation. A supplement to fill the cooling agent tank without having to dismount the supply system is available. Alternatively, an efficient, economical closed-loop intracooler is available, which operates between -85°C and 600°C. Compressed air cooling via compressor or central supply in the laboratory serves for tests above room temperature.

A wide range of crucibles (pans) (aluminum, platinum, stainless steel for pressure applications, etc.) is available for nearly all possible applications and materials.

For the sealing press exchangeable inserts are available, for all cold-weldable aluminum crucibles/pans and stainless steel crucibles (pans) for pressures up to 20 bar (Photo).

The 3in1 Box comes exclusively with the new Concavus® pan by NETZSCH. It offers comprehensive protection during transport and storage, allows for easy withdrawal and features a unique archiving system for the pans following measurement.

The automatic sample changer DSC 204 F1 Phoenix®-ASC is designed for routine measurements. It works day and night, giving you time for other tasks, and provides for optimal use of the DSC 204 F1 Phoenix® (e.g. calibrations on the weekend).

The DSC system with an automatic sample changer can handle up to 192 crucibles/pans evenly distributed on two removable trays. Different types of crucibles/pans are allowed up to 8 mm Ø in and 8 mm in height. A four-needle gripper handles different crucibles by using the appropriate gripping pressure for the chosen pan. For calibration and correction purposes, a fixed strip with additional 12 crucible/pan positions is available. Crucible/pan recognition in flight is available. A crucible/pan and lid data base is linked to the ASC. The sample trays are covered by an automatically controlled cover. After closing the cover, the space above the sample pans is purged by branching gas channels integrated in the cover. The purge gas rate is adapted to opening and closing the cover. For this purpose, a further purge gas inlet is available only for the use with the ASC. A “remove lid” function is integrated to cover the sample while waiting its turn to be inserted into the DSC cell. Alternatively, a piercing device is optionally available for piercing the lid prior to measurement. The DSC system has a refuse bin for disposing lids and non-reusable pans. It is possible to archive the micro-plate trays (storage sample). For better identification, the plates have a serial number and 2D code. The tray identification feature is linked to a crucible/lid data base.The same ASC can also be used with our new TG 209 F1 Libra®.

To analyze the UV curing of light-sensitive resin systems, adhesives, paints, coatings and dental masses, the DSC 204 F1 Phoenix® can be configured as a Photo-DSC. Standard lamps by a variety of manufacturers can serve as a source of radiation.

For adhesives, we recommend the DELOLUX04 high-pressure mercury lamp (315-500 nm), which can easily be triggered using Proteus® software.
For paints and inks, we recommend the DSC be equipped with an OmniCure® S2000 lamp. The measurement parameters (e.g. exposure time and intensity) can be programmed and monitored by segment via the Proteus® software.
The luminous intensity of the high-pressure mercury lamp can be calibrated with the Radiometer R2000.

Your device is too small to show this table.
Standard Crucibles (Pans) Made of Aluminum for the DSC 204 F1 Phoenix®
Material
(Purity)
Temperature
Range
Consisting ofDimension/
Volume
RemarksOrder Number
Al (99.5)Max. 600°CCrucible + lidø 6 mm; 25/40 μl*Set of 100, cold weldable**6.239.2-64.5.00
Al (99.5)Max. 600°CCrucible + lidø 6 mm; 25/40 μl*Set of 500, cold weldable**6.239.2-64.51.00
Al (99.5)Max. 600°CCrucible + lid with
laser-cut hole (50 μm)
ø 6 mm; 40 μlSet of 100, cold weldable**6.239.2-64.8.00
Al (99.5)Max. 600°CCrucible + lid with
laser-cut hole (50 μm)
ø 6 mm; 40 μlSet of 500, cold weldable**6.239.2-64.81.00
Al (99.5)Max. 600°CCrucibleø 6.7mm; 85 μlSet of 100 piecesNGB810405
Al (99.5)Max. 600°CLid100 pieces, for NGB810405NGB810406

* Crucible (pan) volumes of either 25 μl or 40 μl can be achieved by simply reversing the lid.
** One and the same sealing press is capable of sealing all standard Al crucibles (pans); order no. 6.240.10-80.0.00A.

Your device is too small to show this table.
Crucibles (Pans) for General Applications – DSC 204 F1 Phoenix®
Material
(Purity)
Temperature
Range
Consisting ofDimension/
Volume
RemarksOrder Number
Al2O3 (99.7)Max. 1700°CCrucibleø 6.8 mm/85 μlGB399972
Al2O3 (99.7)Max. 1700°CLidFor GB399972GB399973
Fused silicaMax. 1000°CCrucibleø 6.7 mm/85 μlGB399974
Fused silicaMax. 1000°CLidFor GB399974GB399975
Pt/Rh (80/20)Max. 1700°CCrucibleø 6.8 mm/85 μlGB399205
Pt/Rh (80/20)Max. 1700°CCrucibleø 6.8 mm/190 μlNGB801556
Pt/Rh (80/20)Max. 1700°CLidFor GB399205 and NGB801556GB399860
Gold (99.9)Max. 900°CCrucible + Lidø 6.7 mm/85 μl6.225.6-93.3.00
SilverMax. 750°CCrucible + Lidø 6.7 mm/85 μl6.225.6-93.4.00

Literature

Brochure

Brochure

Product brochure: Differential Scanning Calorimetry; Methods, Technique, Applications, 16 pages

Brochure

Application brochure: "Material Characterization, Phase Changes, Thermal Conductivity", 16 pages

Brochure

Application brochure: Research and Development - Substances, Components, Quality Control and Process, Extrusion, Injection Molding, Coating, Quality Assurance - Finished Goods, Material Properties and Main Measuring Methods; 8 pages

Application literature

Application literature
Application literature

Optimization of the temperature profile for curing of a epoxy-based resin by means of DSC data and kinetic analysis using mulitariate non-linear regression

published: NETZSCH-Gerätebau GmbH

Application literature

Phase Transition studies of Cu2As2O7 single crystals by means of XRD and DSC

published: Crystal Growth & Design Vol. 4, No. 6, (2004) 1229

Application literature

Crystal Structure, thermal behavior, vibrational spectroscopy and solid state NMR results of Ag2PO3F

published: Journal of Inorganic Chemistry, 46 (2007) 801

Application literature
Application literature

Investigation of different LDPE LLDPE and HDPE blends with DSC and DMA regarding the performance in terms of morphology and mobility of the amorphous phase

published: Polymer Degradation and Stability 93 (2008) 43

Application literature
Application literature
Application literature

Determination of thermal diffusivity, volumetric expansion and specific heat of NPL reference material inconel 600

published: High Temperatures - High Pressures, 35/36 (2003/2007) 621

Application literature

Determination of the degree of crystallinity of specific grades of polyolefines by means of DMA, DSC WAXS and density

published: J. Polym. Res. 15 (2008) 83

Application literature

Determination of melting behavior, crystallinity and thermal stability of polyolefines by means of DSC

published: Kunststoffe 10/2008

Application literature
Application literature

Oxidative-induction time (OIT) measurements allow for characterization of the long-term stability of hydrocarbons. For determination of the oxidative stability, standardized test methods by means of DSC (Differential Scanning Calorimetry) are used. OIT tests by means of DSC are internationally recognized. Well-established standards are e.g., ASTM D3895-92, ASTM D6186, EN 728 und ISO 11357-6.

Application literature

Poster

Poster

Our poster “Thermal Properties of the Elements” shows the following values at a glance: melting point, boiling point, specific heat, heat of fusion, thermal expansion coefficient, thermal conductivity and density.

Poster

Our poster “Thermal Properties of Polymers” offers the following characteristic values at a glance: Glass transition temperature, heat of fusion and melting temperature, decomposition temperature, E modulus, coefficient of thermal expansion, specific heat capacity, thermal conductivity and density.

Textbook

Press

White Paper

Accessories Catalogue

Accessories Catalogue

Crucibles, Sensors, Sample Carriers, Calibration Kits for DSC, TGA and STA Systems

Digital Media

Video

Video

Designed for routine measurements, it works day and night, giving you time for other tasks, and provides for optimal use of the DSC 204 F1 Phoenix® (e.g. calibrations on the weekend).

(length: 00:27 min)

Video

The versatile premium DSC 204 F1 Phoenix® is based on a unique concept: all of the essential functional modules are integrated into the instrument, and any desired extensions can be added on location at any time. (length: 02:19 min)

NETZSCH DSC 204 F1 Phoenix®
NETZSCH DSC 204 F1 ASC.flv
NETZSCH DSC 204 F1 Phoenix for Differential Scanning Calorimetry
Overview of AutoEvaluation (DSC, TGA) and *Identify (DSC, TGA, DIL/TMA, cp)

Webcast

Quality improvements in plastic parts through powerful thermo-analytical methods
Characterization of Food by Means of Thermal Analysis
How Methods of Thermal Analysis Support the Development, Production and Quality Assurance of Composites
DSC Analysis of Polymers for Experienced Users
Key Thermal Analysis Techniques for Battery Material Development and Testing
DSC Analysis of Polymers – Basics & Applications
Thermal Analysis Methods – Perfect for the Characterization of Wood-Plastic Composites
Thermal Analysis Goes Green – Applications in the Field of Renewable Energies

You may also be interested in:

DSC 204 HP Phoenix® – High-pressure DSC

DSC measurements with selectable gas pressure allow additional possibilities for the study of chemical reactions, vaporization processes (ASTM E1782), adsorption and desorption, and aging studies (ASTM E1858, ASTM D6186, ASTM E2009, ASTM D5483). The DSC 204 HP Phoenix®, Differential Scanning Calorimeter, is ideally configured for these applications.

DSC 214 Polyma
  • High Performance
  • Smart Measurement
  • AutoEvaluation
  • Innovative Identification
  • Easy Sample Preparation
  • Unique Concavus® Pan
DSC 3500 Sirius

The DSC 3500 Sirius combines compactness, robustness and ease-of-use with high detection sensitivity, thus offering optimal conditions for quality assurance and failure analysis in areas such as foods, cosmetics, polymers, technical textiles, and organic and inorganic materials.