Ribavirin — Detection of the melting range

As per UPS <892>, the melting point of a crystalline substance is correlated with the extrapolated onset temperature of the DSC curve, called melt onset temperature. For ribavirin, this value – at 168.0°C – is in good agreement with the melting range (168.1 to 169.8°C, highlighted in red) which was determined for this pharmaceutical ingredient by means of the capillary method (e.g., in line with USB <741>). In the capillary method, detection of the melting range (interval between the beginning of melting and the complete transition into the liquid phase) takes place visually.  

The capillary method was carried out at heating rates of 1 to 1.5 K/min.
The conditions for the DSC measurements were: sample weight: 1.02 mg, heating rate: 10 K/min, nitrogen atmosphere, Al crucibles.

Related Products:

DSC 204 F1 Nevio

Precise and Reliable DSC Results for Pharmaceuticals, Foods and Cosmetics by means of the DSC 204 F1 Nevio, the premium differential scanning calorimeter for research and development, material characterization and contract testing.

DSC 214 Nevio

Studying Phase Transformations of Pharmaceuticals, Foods and Cosmetics by means of the DSC 214 Nevio ‒ the fastest heat-flux DSC instrument on the market.