Transfer line coupling

The transfer line coupling of the TGA or STA thermal analyzer via a heated coupling adapter and a heated transfer line transports the gases evolved to the Bruker INVENIO. The heated gas cell can be located in an external module or connected to the INVENIO via the right exit, or the gas cell may be situated at the internal position. An external gas cell allows the operator access to the FT-IR spectrometer’s great flexibility and makes it possible to add on accessories such as an ATR unit or a microscope.

The gas transfer line is characterized by a low volume and a short capillary length, thus minimizing dilution effects and leading to fast response times. To prevent cold spots and thus condensation of the gases evolved, the entire gas path is heated.

Key Technical Data

(subject to change)

  • Wave number range:
    FT-IR: 8000 cm-1 to 340 cm-1
    Coupling: 4400 cm-1 to 600 cm-1
    (double window technique KBr+ZnSe)
  • Resolution:
    better than 0.4 cm-1
  • Temperature of transfer line:
    max. 230°C
  • Furnace adaptor:
    max. 300°C
  • Transfer line material:
    PTFE (exchangeable)
  • Window material gas cell:
    ZnSe + KBr
  • Gas cell volume and length:
    8.7 ml/110 mm
  • Detector:
    DLaTGS or MCT

Accessories

Automatic Sample Changer

The automatic sample changer (ASC) similarly handles routine measurements in quality control as well as requirements for research and development. It works around the clock to free your time for other challenges and allows you to make optimal use of the coupling even during the weekend. Of course, each sample can be assigned a different measurement and evaluation program. Easy-to-understand input fields lead you through the programming of a series of measurements. Unplanned analyses can also be inserted into a pre-programmed series of measurements already in progress.

Literature

Прикладная литература

Прикладная литература
Прикладная литература

Several standards such as NaHCO3 and CuSO4 * 5H2O are investigated by evolved gas analysis (TG-FTIR and TG-MS) in order to identify and quanitfy the relreased amount of gases

JTAC, Vol 83 (2006) 2, 321

In order to download, the complete paper, please visit Journal of Thermal Analysis and Calorimetry: springerlink.metapress.com/content/f87806577257t86v/

Прикладная литература

Several standards such as NaHCO3 and CuSO4 * 5H2O are investigated by evolved gas analysis (TG-FTIR and TG-MS) in order to identify and quanitfy the relreased amount of gases.

 

published: Thermochim. Acta, 440 (2006) 81

Прикладная литература

Characterization of Dust Samples in Occupational and Environmental Medicine

published: NETZSCH Onset 4/2008

Брошюра

You may also be interested in:

Direct Coupling PERSEUS® STA/TGA
STA 449 F3 Jupiter®

The STA 449 F3 Jupiter® combines the advantages of a high sensitive thermobalance and a true Differential Scanning Calorimeter. Various furnaces and TG, TGA-DTA and TGA-DSC sensors can be used, whereby the system can be easily optimized for the most versatile applications.

QMS 403 Aëolos Quadro - Quadrupole Mass Spectrometer Coupling

The QMS 403 Aëolos Quadro is a very compact mass spectrometer with a heated capillary inlet system, specially designed for routine analysis of volatile decomposition products of thermal analysis (instrument for EGA: evolved gas analysis).

TG 209 F1 Libra®

Flexibility with the fittings of this premium instrument for thermogravimetry allow it to cover almost any imaginable application in the range up to 1100°C. The system has many outstanding features: BeFlat® function allows or immediate measurement upon switching on the unit, the system’s heating rates of up to 200 K/min allow for faster results, while its corrosion-resistant ceramic furnace withstands even fluorine- or chlorine-containing polymers, c-DTA® function for the detection of exothermal and endothermal reactions.

TGA/STA-GC-MS Coupling

NETZSCH, in collaboration with JAS (Joint Analytical Systems), has forged a breakthrough improvement in evolved gas analysis (EGA). What sets the new coupling system apart is the fact that the start of the measurement is event-driven. This enables temperature-correlated detection of the substances released, which in turn allows for direct correlation with mass loss steps.

STA 409 CD-QMS 403/5 SKIMMER - Coupling System

Highest detection sensitivity through direct coupling of the STA 409 CD with a quadrupole mass spectrometer (QMS) - evolved gas analysis (EGA) method.